User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 1-39
Sweep Measurements
Making a Measurement
Step 5. Set up the measurement parameters
You set the output parameters on the MEASURE: SWEEP SETUP screen.
1. Press Meas front-panel key. The MEASURE: SWEEP SETUP screen appears.
In the screen, the upper left area defines the VAR1 information, and the upper
right area defines the VAR1’ information. See Figure 1-8.
2. Set the VAR1 information as shown below:
Drain voltage sweeps from 0 V to 2 V with 20 mV step. The current compliance
is set to 100 mA.
3. Set the VAR1’ information as shown below:
To force the same voltage to the drain and gate, set RATIO = 1 and OFFSET = 0.
Because VAR1’ is defined as follows:
(VAR1’ output) = RATIO × (VAR1 output) + OFFSET
on Front Panel on Keyboard
To move the pointer, use arrow keys of
MARKER/CURSOR area.
use arrow keys.
To set "SINGLE" in
SWEEP MODE field,
select SINGLE secondary
softkey.
press
Shift-F1 keys.
To set "LINEAR" in
LIN/LOG field,
select LINEAR secondary
softkey.
press
Shift-F1 keys.
To enter "2.000 V"
in STOP field,
press
2, then Enter.type 2,
then press
Enter.
To enter "20.00 mV"
in STEP field,
press
2, 0, m, then Enter.type 20m,
then press
Enter.
on Front Panel on Keyboard
To enter "0.000 V" in
OFFSET field,
press
0, then Enter.type 0, then press Enter.
To enter "1.000" in
RATIO field,
press
1, then Enter.type 1, then press Enter.