User`s guide
1-38 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sweep Measurements
Making a Measurement
Figure 1-7 User Function Definition Screen
Figure 1-7 defines the following two user functions:
SQID = √ Id
DSQID = ∂SQID / ∂Vg = ∂√ Id / ∂Vg
Where, Id is drain current and Vg is gate voltage.
B
CHANNELS: USER FUNCTION DEFINITION 01JAN29 10:53AM
S
CHANNEL
DEF
USER
FCTN
USER
VAR
E5250A
PROP
PREV
PAGE
NEXT
PAGE
DELETE
ROW
SQID
Enter User Function Name. (max 6 chars.)
*USER FUNCTION
NAME UNIT DEFINITION
SQID SQRT(ID)
DSQID DIFF(SQID,VG)