User`s guide

Agilent 4155C/4156C Users Guide Vol.2, Edition 6 1-35
Sweep Measurements
Making a Measurement
Step 3. Define the channel assignments
You set the connection information on the CHANNELS: CHANNEL DEFINITION
screen.
1. Turn on the 4155C/4156C. Self-test starts.
2. After self-test is finished, make sure that CHANNELS: CHANNEL
DEFINITION screen appears on the screen of the 4155C/4156C. If not, press
Chan front-panel key.
where, softkeys located bottom of screen are the primary softkeys, and softkeys
located right side of screen are the secondary softkeys.
3. Make sure that SWEEP is displayed in the MEASUREMENT MODE field. If
not, select SWEEP secondary softkey in the MEASUREMENT MODE field.
4. Set the connection information in the CHANNELS area as shown in the
following table. For the actual screen, see Figure 1-6.
on Front Panel on Keyboard
To move the pointer, use arrow keys of
MARKER/CURSOR area.
use arrow keys
To move the cursor to
edit in display area,
use arrow keys of
Edit area.
use
Backspace key.
To enter "VS" in
VNAME field,
press
. (period) and + keys,
then
Enter.
type VS, then press
Enter.
To enter "IS" in
INAME field,
press
* and + keys,
then
Enter.
type IS, then press
Enter.
To set "V" in
MODE field,
select V secondary softkey. press
Shift-F1 keys.
To set "VAR1" in
FCTN field,
select VAR1' secondary
softkey.
press
Shift-F4 keys.
To set "VAR1" in
FCTN field,
select VAR1 secondary
softkey.
press
Shift-F2 keys.
To disable a unit, select DISABLE UNIT
secondary softkey.
press
Shift-F7 keys.