User`s guide

8-32 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Support Functions
Switching Matrix Control
To Control Functions
To control the E5250A functions, use the CHANNELS: E5250A PROPERTIES
screen. Press the
Chan front panel key, then select the E5250A PROP primary
softkey. The E5250A PROPERTIES screen appears.
Step 1. To set GPIB address
Enter the GPIB address of the E5250A into the GPIB ADDRESS field.
Step 2. To set control mode
In the CONTROL field, select the ON softkey.
Then the 4155C/4156C sends a query for the present setup of the E5250A, and
displays it. In the E5250A control ON status:
The E5250A SETUP primary softkey appears. Selecting the softkey displays the
CHANNELS: E5250A CONNECTION SETUP screen.
You can access the E5250A setup (extension: MAT) file using the filer function.
If you disable the E5250A control, select the OFF softkey.
B
CHANNELS: E5250A PROPERTIES 01JAN15 05:04PM
S
CHANNEL
DEF
USER
FCTN
USER
VAR
E5250A
PROP
22
Enter GPIB address of E5250A (0 to 30).
*E5250A SETUP
GPIB ADDRESS 22
CONTROL OFF
CARD TYPE E5252A
CONFIG MODE NORMAL
*MATRIX CONNECTION MODE
PORT BIAS COUPLE PORT STATUS CONN CONN
FCTN PORT 13579RULESEQ
CARD1 NO FCTN ---- --- --- --- --- --- FREE BBM
CARD2 NO FCTN ---- --- --- --- --- --- FREE BBM
CARD3 NO FCTN ---- --- --- --- --- --- FREE BBM
CARD4 NO FCTN ---- --- --- --- --- --- FREE BBM
*E5250A INPUT CONNECTION
INPUT 1 SMU1 INPUT 5 SMU5
INPUT 2 SMU2 INPUT 6 SMU6
INPUT 3 SMU3 INPUT 7 VSU1
INPUT 4 SMU4 INPUT 8 VSU2
INPUT 9 VMU1
INPUT10VMU2