User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 8-27
Support Functions
SMU/PG Selector Control
To Use Selector
1. Press Stress key in the PAGE CONTROL key group.
2. Select CHANNEL DEF primary softkey.
3. In the MEASURE field of the SMU/PG SELECTOR area, select one of the
following softkeys. This field sets the switching status of selector in the
measurement state.
• SMU secondary softkey to connect SMU to DUT
• PGU secondary softkey to connect PGU to DUT
• OPEN secondary softkey to disconnect SMU, PGU, and DUT
• PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch
4. In the STRESS field of the SMU/PG SELECTOR area, select one of the
following softkeys. This field sets the switching status of selector in the stress
force state.
• SMU secondary softkey to connect SMU to DUT
• PGU secondary softkey to connect PGU to DUT
• OPEN secondary softkey to disconnect SMU, PGU, and DUT
• PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch