User`s guide
8-18 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Support Functions
R-BOX Control
R-BOX Control
Agilent 16441A R-Box must be used for applications which need to connect a series
resistor between SMU and DUT. For example, the R-Box is effective for the DUT
protection. If sudden voltage change occurs at DUT, excessive current flows to the
DUT, and it may damage the DUT without the R-Box. In other case, you may want
to measure negative resistance characteristics. This application needs series resistor
because SMUs cannot measure negative resistance.
The 4155C/4156C automatically compensates for voltage drop of the series
resistance value. So, the GRAPH/LIST screen group show the compensated data.
For details of the 16441A R-Box, refer to Agilent 16441A R-Box User's Guide.
Resistance Value
The 16441A R-Box provides two sets of the series resistors. So you can connect two
SMU channels to the R-Box. Resistance values of resistors are listed below:
•1 MΩ
• 100 kΩ
•10 kΩ
•0 Ω
You can select the resistance values on the 4155C/4156C setup screen. The LEDs on
the 16441A R-Box indicate the present resistance value.
Connections
Following table is the parts list of cables for connecting the 16441A R-Box.
Agilent Model or Part Number Description
04155-61610 Control Cable 1.5 m
04155-61609 Control Cable 3.0 m
04155-61605 Triaxial Cable 0.4 m
16493K-001 Kelvin Triaxial Cable 1.5 m
16493K-002 Kelvin Triaxial Cable 3.0 m