8-6 Agilent 4155C/4156C Users Guide Vol.2, Edition 6
Support Functions
User Function and User Variable
User Variable
A user variable is a data variable that is a numeric list, which is passed via GPIB
commands of PAGE:CHANnels:UVARiable and TRACe|DATA subsystems from
an external computer or the Internal IBASIC. For information about the
PAGE:CHANnels:UVARiable and TRACe|DATA subsystems, refer to GPIB
Command Reference.
You can perform calculations between measurement results and the numeric list, or
plot the numeric list on the GRAPH/LIST: GRAPHICS screen.
To define a user variable, you define a name and an data size on the CHANNELS:
USER VARIABLE DEFINITION screen. If desired, you can define a unit, such as
ms. You can define up to 6 user variables. See example below.
User variable name must start with alphabet character and can consist of
maximum six alphanumeric characters. Name must be unique. Name is case
sensitive. For example, Gm is different from gm.
Unit name is optional. Length: 1 to 6 characters. Valid characters: any
characters.
Example
To define a user variable for mutual conductance gm of an FET, define gm on the
CHANNELS: USER VARIABLE DEFINITION screen as follows:
NAME UNIT SIZE
gm S 101