User`s guide
7-54 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Measurement Units and Functions
Measurement Sequence
Measurement Sequence
The 4155C/4156C performs measurements for the variables or user functions set to
the following entry fields on the DISPLAY: DISPLAY SETUP screen. Then
measurements will be performed in the following order. For the user functions,
measurements will be performed for the variables defined in the functions.
1. NAME for X axis of the GRAPHICS table
2. NAME for Y1 axis of the GRAPHICS table
3. NAME for Y2 axis of the GRAPHICS table
4. NAME for No. 1 of the LIST table
5. NAME for No. 2 of the LIST table
6. NAME for No. 3 of the LIST table
7. NAME for No. 4 of the LIST table
8. NAME for No. 5 of the LIST table
9. NAME for No. 6 of the LIST table
10. NAME for No. 7 of the LIST table
11. NAME for No. 8 of the LIST table
12. Upper field of DATA VARIABLES
13. Lower field of DATA VARIABLES
Exceptions If a variable is set to multiple fields or user functions, measurement will be
performed only for the first entry.
If a variable not set to the fields is defined in a user function not set to the fields, the
4155C/4156C performs measurement for the variable after the measurement for the
variable set to the lower field of DATA VARIABLES.
For the sampling measurements that use the stop condition, the 4155C/4156C
performs measurement for the variable set to the following field before the
measurement for the variable set to the NAME for X axis.
• NAME, STOP CONDITION table, MEASURE: SAMPLING SETUP screen