Measurement Units and Functions
QSCV Zero Offset Cancel
Table 7-27 Measurement Condition for the Offset Measurement
Parameter for the
offset measurement
Internal setup (not the setup on the screen)
Start voltage 0 V
Stop voltage Minimum value of the following values:
|setup value of the QSCV MEAS VOLTAGE field| × 2
|setup value of the START field|
|setup value of the STOP field|
or 5 V (if the all listed above is greater than 5 V)
or 10 mV (if the all listed above is less than 10 mV)
Hold time setup value of the HOLD TIME field
Delay time 0 second
Integration time for
capacitance
measurement
a
a. If the integration time was too short for the offset measurement, the value
is automatically changed to the setup value of the QSCV field, and the off-
set measurement is executed again.
0.5 sec + |setup value of the STOP field|
1/4
sec
for the 10 pA or 100 pA range
80 msec for the 1 nA or 10 nA range
Integration time for
leakage current
measurement
b
b. The leakage current data is just used for the offset data calculation.
setup value of the LEAK field