User`s guide

Agilent 4155C/4156C Users Guide Vol.2, Edition 6 7-43
Measurement Units and Functions
QSCV Zero Offset Cancel
QSCV Zero Offset Cancel
This function is available for the quasi-static CV measurements. This function
enables you to minimize measurement error (offset) caused by stray capacitance of
cables, prober, and so on.
To Measure Offset Data
Measure the offset data, as follows:
1. Set the measurement conditions on the MEASURE: QSCV SETUP and
MEASURE: QSCV MEASURE SETUP screen.
The QSCV measurement setup must be completed before executing the offset
measurement. The offset measurement uses setups on the screen. See Table
7-27.
2. Open the measurement terminals at the cable end of the device side.
3. Select the ON softkey in the ZERO CANCEL field on the MEASURE: QSCV
MEASURE SETUP screen.
4. Press the green key and
Stop front-panel key in this order.
The measurement unit executes a one-point capacitance measurement between
the start voltage and the stop voltage shown in Table 7-27. And the measured
offset data is displayed in the field right next to the ZERO CANCEL field. The
initial offset data is 0.
NOTE Invalid Offset Data
If one of the following conditions is detected during the offset measurement, an
error message is displayed, and the offset data measurement is aborted.
Integration time is too short for the offset measurement
Leakage current is more than 70% of the measurement range
When this happens, change the measurement conditions and execute the offset
measurement again.