Measurement Units and Functions
Measurement Ranging Mode
Auto Ranging
The monitor unit automatically searches for and measures at the range that provides
the highest resolution as follows:
V measurement
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 110 % of the range, then the unit performs the
measurement.
I measurement
•1 A to 1 μA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
If the measurement value is less than 1 % of the present range and the present
range is 100 μA or higher range, the range changes down two ranges instead of
one range.
100 nA to 100 pA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
10 pA
The unit changes to the next higher range when the measurement value exceeds
104 % of the present range.
The auto ranging operation can be expanded for the current measurement. See
“Enhanced Auto Ranging for Current Measurement” on page 7-33.