User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 7-25
Measurement Units and Functions
Compliance
Table 7-18 V Compliance Resolution
Table 7-19 I Compliance Setting Range
HPSMU 1 nA to 10 mA 0 to 200 V
100 mA (0 ≤ |I| ≤ 50 mA) 0 to 200 V
100 mA (50 mA < |I| ≤ 115 mA) 0 to 100 V
1 A (0 ≤ |I| ≤ 50 mA) 0 to 200 V
1 A (50 mA < |I| ≤ 125 mA) 0 to 100 V
1 A (125 mA < |I| ≤ 500 mA) 0 to 40 V
1 A (500 mA < |I| ≤ 1 A) 0 to 20 V
Unit V Compliance Resolution
HRSMU 0 V ≤ |V| ≤ 2 V 100 μV
MPSMU 2 V < |V| ≤ 20 V 1 mV
HPSMU 20 V < |V| ≤ 40 V 2 mV
40 V < |V| ≤ 100 V 5 mV
100 V < |V| ≤ 200 V 10 mV
Unit Output Range I Compliance Setting Range
HRSMU 2 V 100 fA to 100 mA
20 V 100 fA to 100 mA
40 V 100 fA to 50 mA
100 V 100 fA to 20 mA
MPSMU 2 V 1 pA to 100 mA
20 V 1 pA to 100 mA
40 V 1 pA to 50 mA
100 V 1 pA to 20 mA
Unit Output Range
V Compliance
Setting Range