User`s guide

7-22 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Measurement Units and Functions
Measurement Units
Leading Edge and Trailing Edge Transition Time
The leading-edge and trailing-edge transition times have five setting ranges as
shown in Table 7-16.
Table 7-16 Ranges and Resolutions of Leading and Trailing Transition Time
where,
leading-edge transition time < pulse width × 0.8
trailing-edge transition time < (pulse periodpulse width) × 0.8
Leading and trailing-edge transition times for a PGU must be in the range
specified by a Range.
Trigger Output
PGUs output trigger signal to synchronize with external pulse generators. If an
41501A/B has PGUs, you can see a trigger output terminal on its rear panel. For
details of trigger functions, see Chapter 8.
Range Leading and Trailing Transition Time Resolution
1 100 ns to 1000 ns 1 ns
20.50 μs to 10.00 μs 10 ns
35.0 μs to 100.0 μs100 ns
450 μs to 1000 μs1 μs
5 0.5 ms to 10.00 ms 10 μs