7-16 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Measurement Units and Functions
Measurement Units
• Voltage Compliance
If you use SMU as pulse current source, you can set voltage compliance as
follows:
• When |I| ≤ 10 μA, voltage compliance must be 2 V or less.
• When |I| > 10 μA, voltage compliance ranges are same as in tables on
previous pages.
where, meaning of I depends on the pulse output mode. See below:
If SMU is pulsed constant source peak or base value, whichever has larger
absolute value.
If SMU is pulsed sweep source start or stop value, whichever has larger
absolute value.