Measurement Units
HRSMU - High Resolution SMU
Figure 7-3 HRSMU Output and Measurement Ranges
Table 7-1 HRSMU Output Voltage Ranges and Resolutions
Range Output Value
Output
Resolution
Current
Compliance
Range
2 V 0 |V| 2 V 100 μV ±100 mA
20 V 0 |V| 20 V 1 mV ±100 mA
40 V 0 |V| 40 V 2 mV ±50 mA