User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 1-5
Sweep Measurements
Measurement Functions
Figure 1-1 Basic Sweep Measurement
To set up basic sweep measurement, select VAR1 function for desired SMU or VSU
on CHANNELS: CHANNEL DEFINITION screen.
Parameters
Also, specify the following parameters for VAR1 on MEASURE: SWEEP SETUP
screen.
Parameter Description
sweep mode Single or double sweep.
linear/log Linear or logarithmic sweep. For logarithmic sweep, select the
number steps in one decade as follows:
LOG10 10 steps in one decade.
LOG25 25 steps in one decade.
LOG50 50 steps in one decade.