User`s guide

5-8 Agilent 4155C/4156C Users Guide Vol.2, Edition 6
Stress Force
Stress Function
NOTE Pulse Waveform when Stress Stops
When you set the duration mode or press the
Stop front-panel key, be aware that
stress force may stop during the pulse peak output as shown in the following figure:
Sequence for returning to 0 V (stress force state to the idle state)
When the state changes from the stress state to the idle state, the outputs of the
channels are returned to 0 V in opposite order that forcing occurred.
Delay time of PGUs
When PGUs are set to VPULSE (ac source), you can set a delay time as follows:
If PGU is set to SYNC, the PGU waits the delay time (after the stress start trigger
is received), then starts to force ac stress.
If PGU is set to NSYNC, the PGU waits the delay time (after stress force state
starts), then starts pulse output.
Example
Figure 5-1 shows an example of output sequence when forcing stress.
Figure 5-1 assumes the output sequence is set on the MEASURE: OUTPUT
SEQUENCE screen as follows.
1. PGU1
2. SMU1
3. PGU2
4. SMU2
5. SMU3