User`s guide
4-40 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Quasi-static C-V Measurements
Considering Measurement Accuracy
Figure 4-12 Calculation Example of Measurement Accuracy: HRSMU
Conditions:
Measurement Range: 1 nA
Output Range: 40 V
Integration Time: 2, 1, 0.5, 0.3, 0.1, 0.05, 0.03 s corresponding to the lines from top
to bottom
Equivalent Parallel Resistance of DUT: 1 G ohm
Guard Capacitance of Measurement Path: 200 pF
0
2
4
6
8
10
0.01 0.1 1 10
QSCV Measurement Voltage (V)