User`s guide

4-24 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Quasi-static C-V Measurements
Making a Measurement
Step 6. Set up the results display
Set the results display information on the DISPLAY: DISPLAY SETUP screen.
1. Press the Display front-panel key. The DISPLAY: DISPLAY SETUP screen
appears.
2. Move the field pointer to the DISPLAY MODE field. Then select the
GRAPHICS softkey to set the results display mode to the graph mode.
3. Set the X and Y1 axes information. The following values are just an example.
This setup is to assign the X axis to V1, and the Y1 axis to CAP01.
X-axis Y1-axis
NAME V1 CAP01
SCALE LINEAR LINEAR
MIN 3 V 36 pF
MAX 3 V 110 pF
B
DISPLAY: DISPLAY SETUP
S
DISPLAY
SETUP
ANLYSIS
SETUP
PREV
PAGE
NEXT
PAGE
GRAPH-
ICS
LIST
GRAPHICS
Select Display Mode with softkey or rotary knob.
*DISPLAY MODE
GRAPHICS
*GRAPHICS
Xaxis Y1axis Y2axis
NAME V1 CAP01
SCALE LINEAR LINEAR
MIN -3.0000000 V 36.00000pF
MAX 3.0000000 V 110.00000pF
*GRID *LINE PARAMETER
ON ON
*DATA VARIABLES *DATA DISPLAY RESOLUTION
EXTEND