User`s guide

Agilent 4155C/4156C Users Guide Vol.2, Edition 6 4-23
Quasi-static C-V Measurements
Making a Measurement
4. Enable or disable the leakage current compensation function using the LEAK
COMPENSATION field.
To enable the function, select the ON softkey.
To disable the function, select the OFF softkey.
5. Enable or disable the zero offset cancel function using the ZERO CANCEL
field. To disable the offset cancel function, select the OFF softkey.
The following procedure enables the function.
a. Complete the QSCV measurement setup on the MEASURE: QSCV SETUP
and MEASURE: QSCV MEASURE SETUP screens.
b. Remove the DUT from the socket module to open the measurement
terminals.
c. Press the green front-panel key and the
Stop front-panel key in this order to
execute the offset measurement. And wait for the measurement completion.
d. Select the ON softkey in the ZERO CANCEL field.
e. Mount the DUT on the socket module.
B
MEASURE: QSCV MEASURE SETUP
S
QSCV
SETUP
MEASURE
SETUP
OUTPUT
SEQ
PREV
PAGE
NEXT
PAGE
SMU1:HR
SMU2:HR
SMU1:HR
Select Measurement Unit with softkey.
*MEASUREMENT UNIT
UNIT FCTN RANGE CNAME INAME
SMU1:HR VAR1 1nA CAP01 LEAK01
*INTEG TIME
TIME NPLC
QSCV 100ms 5
LEAK 100ms 5
*LEAK COMPENSATION
ON
*ZERO CANCEL
ON 5.416pF