User`s guide

Agilent 4155C/4156C Users Guide Vol.2, Edition 6 4-19
Quasi-static C-V Measurements
Making a Measurement
Step 3. Define the channel assignments
Set the connection information on the CHANNELS: CHANNEL DEFINITION
screen.
1. If the 4155C/4156C has been turned off, turn it on and wait until the self-test is
completed.
2. Make sure that the CHANNELS: CHANNEL DEFINITION screen appears on
the screen of the 4155C/4156C. If not, press the
Chan front-panel key.
3. Move the field pointer to the MEASUREMENT MODE field, then select the
QSCV softkey to set the 4155C/4156C to QSCV measurement mode.
4. Move the field pointer to the CHANNELS table, and set SMU1 and SMU2 as
shown below.
Only the SMU1 and SMU2 are used for this example measurement. Disable
other units by selecting the DELETE ROW softkey.
UNIT VNAME INAME MODE FCTN
SMU1 V1 I1 V VAR1
SMU2 V2 I2 COMMON CONST