User`s guide

Agilent 4155C/4156C Users Guide Vol.2, Edition 6 4-3
Quasi-static C-V Measurements
Measurement Functions
Measurement Functions
For quasi-static C-V (QSCV) measurements, the sweep source channel performs
linear staircase sweep output of voltage, while the monitor channel measures
capacitance for each sweep step.
Available Units
The QSCV measurement uses a voltage sweep source (VAR1), and a voltage
constant source. Also, optional constant sources may be used. Units, mode, and
function available are shown below:
Only a SMU can be used for the capacitance measurement. Specify the
measurement unit on the MEASURE: QSCV MEASURE SETUP screen.
Capacitance Data Calculation
At each sweep step, voltage and current are measured and the capacitance data is
calculated by using the following formula.
C = I × cinteg /(V - V
0
)
where, cinteg is the integration time of the capacitance measurement. I, V, and V
0
are the measurement data at each sweep step. Also, if the leakage current
compensation is enabled, the capacitance data is calculated by the next formula.
C = (I - Leak) × cinteg /(V - V
0
)
Leak is the leakage current and is given by the following formula.
Leak = IL + τ × (IL
0
- IL)/(2 × cinteg)
where, τ is the time for electrical charge.
Unit Mode Function
SMU V VAR1
V, I, COMMON CONST
VSU V CONST
PGU V CONST
GNDU COMMON CONST