3-40 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sampling Measurements
Making a Measurement
Step 5. Set up the results display
You set the results display information on the DISPLAY: DISPLAY SETUP screen.
1. Press Display front-panel key. The DISPLAY: DISPLAY SETUP screen appears.
2. Move the field pointer to the DISPLAY MODE field. Then select GRAPHICS
softkey to set the results display mode to the graph mode.
3. Set the X and Y1 axes information as shown below:
This setup is to assign X-axis to time, and Y1-axis to VG.
4. Move the field pointer to the DATA VARIABLES field. Then select the
@INDEX softkey to display the data index above graph in the results display
screen (GRAPH/LIST: GRAPHICS screen).
X-axis Y1-axis
NAME @TIME VG
SCALE LINEAR LINEAR
MIN 0 s 0 V
MAX 200 ms 1 V
B
DISPLAY: DISPLAY SETUP 00DEC07 02:18PM
DISPLAY
SETUP
ANLYSIS
SETUP
PREV
PAGE
NEXT
PAGE
GRAPH-
ICS
LIST
GRAPHICS
Select Display Mode with softkey or rotary knob.
*DISPLAY MODE
GRAPHICS
*GRAPHICS
Xaxis Y1axis Y2axis
NAME @TIME VG
SCALE LINEAR LINEAR
MIN 0.000000000 s 0.000000000 V
MAX 200.00ms 1.000000 V
*GRID
*LINE PARAMETER
ON
ON
*DATA VARIABLES
@INDEX
*DATA DISPLAY RESOLUTION
NORMAL