User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 3-35
Sampling Measurements
Defining Measurement Conditions
To Execute or Stop Measurement
• To execute a measurement, press:
•
Single key in the MEASUREMENT key group for single measurement.
•
Repeat key in the MEASUREMENT key group for repeat measurement.
•
Append key in the MEASUREMENT key group for append measurement.
• To stop a measurement, press
Stop key in the MEASUREMENT key group.
Single, Repeat, and Append Measurement
There are three measurement execution modes as follows:
Single
measurement Clears GRAPHICS or LIST screen, then executes measurement
one time. Measurement results are displayed on GRAPHICS or
LIST screen.
Repeat
measurement Executes measurements continuously. Before each
measurement is executed, the GRAPHICS or LIST screen is
cleared. Most recent measurement results are displayed on
GRAPHICS or LIST screen.
Append
measurement Executes measurement one time. Does not clear GRAPHICS or
LIST screen. That is, measurement results are added to the