User`s guide
3-34 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sampling Measurements
Defining Measurement Conditions
To Display List Results
1. Press Display key in the PAGE CONTROL key group.
2. Select DISPLAY SETUP primary softkey.
3. In the DISPLAY MODE field, select LIST secondary softkey.
4. In the LIST area, select the secondary softkey of the variables for which you
want to list the measurement results.
When the pointer is in the NAME row, the allowable variable names appear in the
secondary softkey area. To set a variable name, select the desired secondary softkey.
The allowable names are names that you already set up on CHANNEL
DEFINITION, USER FUNCTION, and USER VARIABLE screens.
To set up variable to be displayed on the LIST screen
In the DATA VARIABLES fields, select secondary softkey for desired variable. Up
to 2 variables can be displayed on the data variable area which is located below the
list area.
To change resolution of the data to be displayed
In the DATA DISPLAY RESOLUTION field, select NORMAL for the resolution
shown in the specifications, or EXTEND for the ADC full scale resolution. For
example, the data display resolution is 1 fA in NORMAL mode, and 10 aA in
EXTEND mode for 10 pA measurement range.