User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 3-29
Sampling Measurements
Defining Measurement Conditions
To Set up Constant Output
1. Define CONST units as described in “To Define Measurement Units” on page
3-26.
2. Press
Meas key in the PAGE CONTROL key group.
3. Select SAMPLNG SETUP primary softkey.
4. In the SOURCE field of the desired unit in the CONSTANT area, enter the
desired output value.
To modify the UNIT, NAME, and MODE field
Modify the UNIT, NAME, and MODE fields on the CHANNELS: CHANNEL
DEFINITION screen.
To set up compliance value for constant output
Set desired value in the COMPLIANCE field of the CONSTANT table.