Technical data

Agilent B1500A Configuration and Connection Guide, Edition 5 4- 19
Connection and Ordering Examples
Four terminal device
Table 4-25 Target device and application
Table 4-26 Key specification
Table 4-27 Configuration Example
Device Application Note
FET I
d
-V
d
, V
th
, BV
dss
, I
d
leak, I
g
leak
Item Range Note
Maximum I/V range 1 A / 30 V Drain terminal by MCSMU
Maximum I/V range 100 mA / 100 V Gate/Substrate terminal by MPSMU
Minimum I/V resolution 10 pA / 0.2 V Drain terminal by MCSMU
Minimum I/V resolution 10 fA / 0.5 V Gate/Substrate terminal by MPSMU
Minimum pulse width 50 s Drain terminal by MCSMU
I
d
fast sampling 2 s Drain terminal by MCSMU
Model option Description Qty
B1500A Semiconductor Device Analyzer Mainframe 1
B1500A-015/030 1.5 m or 3 m length of bundled cables 1
B1500A-A00 Empty package for custom solution, contains no measurement module 1
B1500A-A11 MPSMU add-on package, adds one MPSMU and two triaxial cables 2
B1500A-A1A MCSMU add-on package, adds one MCSMU, connection box and two
triaxial cables
1
B1500A-A5F Test fixture for packaged device measurement, adds one 16442B (Optional) 1
16494A Triaxial cable 4