Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment • FAST SHIPPING AND DELIVERY • TENS OF THOUSANDS OF IN-STOCK ITEMS • EQUIPMENT DEMOS • HUNDREDS OF MANUFACTURERS SUPPORTED • LEASING/MONTHLY RENTALS • ITAR CERTIFIED SECURE ASSET SOLUTIONS SERVICE CENTER REPAIRS Experienced engineers and technicians on staff at our full-service, in-house repair center WE BUY USED EQUIPMENT Sell your excess, underutilized, and idle used equipment We also offer credit for buy-back
Agilent 16048G/H Test Leads Operation and Service Manual Third Edition Agilent Part No. 16048-90050 January 2001 Printed in: Japan Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd. Component Test PGU-Kobe 1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan © Copyright Agilent Technologies Japan, Ltd.
• DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. • Keep Away From Live Circuits Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected.
Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period corresponding to the individual warranty periods of its component products. Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a period of 90 days. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below.
6 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Contents 1. Installation Guide Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Connecting the 16048G/H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2. Overview Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 8 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
1 Installation Guide 9 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Installation Guide Incoming Inspection Incoming Inspection Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the 16048G/H has been checked mechanically and electrically. The contents of the shipment should be as listed in Table 1-1. If the contents are incomplete, if there is mechanical damage or defect, notify the nearest Agilent Technologies office.
Installation Guide Connecting the 16048G/H Connecting the 16048G/H Follow these steps below to connect the 16048G or 16048H to the 4294A. Step 1. Set the 16048G/H to the test connectors on the front panel of the HP 4294A by gradually coupling the four BNC connectors and fastening screws of the fixture with the test connectors and accessory mounting holes of the instrument until they come to complete contact. Step 2.
Installation Guide Connecting the 16048G/H 12 Chapter 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
2 Overview 13 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Overview Product Overview Product Overview The 16048G/H consists of a direct attachment, 4-terminal pair interface which is equipped with four BNC (f) connectors. These test leads are used to attach user-fabricated test fixtures. Cable length of the 16048G is 1 meter, the 16048H is 2 meter. Figure 2-1 Product Overview 14 Chapter 2 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Overview Functions Functions Figure 2-2 shows name of each part of the 16048G/H and Table 2-2 shows their function. Figure 2-2 16048G/H Parts Table 2-1 16048G/H Function No. NAME FUNCTION 1 Instrument Side Connection Block Connect to the instrument’s UNKNOWN terminal. 2 Connection Block Securing Screw Secures the connection block to the 4294A. 3 Ground Lead Connect to the chassis or grounded part of the measurement target.
Overview Functions Figure 2-3 Cable Assignment 16 Chapter 2 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
3 Operation This chapter describes the proper methods for setting the 4294A, fixture compensation with the 16048G/H. 17 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Operation 4294A Setting 4294A Setting Before you begin your measurement, you should perform the adapter setup in the 4294A. Also refer to the operation manual of the 4294A about the adapter setup. Adapter Setup Connect the 16048G/H to the 4294A and perform the adapter setup described below. For adapter setup, use the 100 Ω Resistor furnished with the 4294A (Agilent P/N 04294-61001). NOTE 1. More than 30 minutes warm-up time is required after turning on the 4294A. 2.
Operation 4294A Setting 7. Press PHASE COMP [-] key to start the phase compensation data measurement. About 1 minutes later, phase compensation data measurement is completed and the softkey label changes to PHASE COMP [DONE]. 8. Connect the Lcur, Lpot, Hcur, Hpot terminal on the 16048G or 16048H to the Lcur, Lpot, Hcur, Hpot terminal of the 100 Ω Resistor respectively. Figure 3-2 Connecting 100 Ω Resister (Load Data Measurement) 9. Press LOAD [-] key to start the load data measurement.
Operation Performing Fixture Compensation Performing Fixture Compensation To compensate stray capacitance and residual impedance, fixture compensation should be performed. When you use the optional BNC Bracket (Agilent P/N 16048-60003) or an user fabricated test fixture, fixture compensation is required. Refer to the 4294A Operation Manual for the fixture compensation procedure. 20 Chapter 3 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Specifications This chapter provides specifications of the 16048G/H Test Leads. 21 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Specifications Specifications Specifications Applicable Instruments LCR meters and Impedance Analyzers with four-terminals Maximum Voltage ± 42V peak max. (AC+DC) Operating Environment temp. -20°C to +150°C (Exclude up to 20 cm from four terminal pair connection block) humidity 15% to 95%RH( @ wet bulb temp. < 40°C) Non Operating temp. -40°C to +70°C Environment. humidity ≤ 90 % RH ( @ wet bulb temp.
5 Service This chapter provides information on servicing and proper maintenance. 23 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Service Maintenance Maintenance An exploded view of the 16048G/H for parts identification is shown in Figure 5-1 to Figure 5-4. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts. To order parts, use the Agilent part numbers listed in Table 5-1 to Table 5-4.
Service Maintenance Table 5-1 Replacable Parts (part 1 of 4) Reference Designator Agilent Part No. Qty.
Service Maintenance Figure 5-2 Replacable Parts (part 2 of 4) Table 5-2 Replacable Parts (part 2 of 4) Reference Designator Agilent Part No. Qty. Description 1 16044-60001 2 CONN BNC-SMB 2 3050-0067 4 WSHR-FL MTLC 3 3050-0789 4 WSHR-FL NM 4 16047-40002 4 INSULATOR 5 1253-0476 2 ADPT BNC-SMB 26 Chapter 5 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Service Maintenance Figure 5-3 Replacable Parts (part 3 of 4) Table 5-3 Replacable Parts (part 3 of 4) Reference Designator Agilent Part No. Qty. Description 1 16047-24021 2 KNOB 2 16048-04021 1 COVER TOP (16048G) 16048-04025 1 COVER TOP (16048H) 3 16047-24026 2 FLANGE 4 0515-0914 4 SCR-MACH M3X0.5 5 0515-0952 4 SCR-MACH M2X0.4 6 16048-00624 1 ANGLE 7 0515-0914 3 SCR-MACH M3X0.5 Chapter 5 Artisan Technology Group - Quality Instrumentation ...
Service Maintenance Table 5-3 Replacable Parts (part 3 of 4) Reference Designator Agilent Part No. Qty. Description 8 16048-04022 1 COVER BOTTOM 9 16048-00625 1 ANGLE Figure 5-4 Replacable Parts (part 4 of 4) Table 5-4 Replacable Parts (part 4 of 4) Reference Designator Agilent Part No. Qty.
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment • FAST SHIPPING AND DELIVERY • TENS OF THOUSANDS OF IN-STOCK ITEMS • EQUIPMENT DEMOS • HUNDREDS OF MANUFACTURERS SUPPORTED • LEASING/MONTHLY RENTALS • ITAR CERTIFIED SECURE ASSET SOLUTIONS SERVICE CENTER REPAIRS Experienced engineers and technicians on staff at our full-service, in-house repair center WE BUY USED EQUIPMENT Sell your excess, underutilized, and idle used equipment We also offer credit for buy-back