Specifications
66
Up to 3 GHz (7 mm) Material
Placing the MUT on the Test Fixture as follows:
Furnished Accessories:
Compensation and Measurement: Short compensation
is required before measurement. Short compensation is
performed by only inserting the MUT holder into the test
fixture. After performing short compensation, the MUT is
inserted into the fixture as shown below.
Description P/N Qty.
Fixture Holder 16454-00601 1
Tweezers 8710-2081 1
Screw, Hex Recess 0515-1050 1
Holder A 16454-25002 1
Holder B 16454-25001 1
Holder C (Without hole) 16454-25004 1
Holder D (With hole) 16454-25003 1
Holder Case 1540-0622 1
Hex Key (for replacing fixtures) 8710-1181 1
Carrying Case 16454-60101 1
Operation and Service Manual 16454-90020 1
Remove the cap of the fixture Place a MUT holder only in the fixture Replace the cap by screwing tightly
Remove the cap of the fixture Place a MUT onto the MUT holder and
insert it into the fixture.
Replace the cap by screwing tightly
Short Compensation
E4991A with 16454A