Specifications
44
16093-65003 Binding Post Test Fixture
Terminal Connector: 7 mm
DUT Connection: 2-Terminal
Electrical Length: 3.4 mm
Dimensions (approx.): 150(W) x 70(H) x 80(D) [mm]
Weight (approx.): 135 g
Additional Error: See figure below
Up to 3 GHz (7 mm) Lead Components
Description: This test fixture is designed for impedance
evaluation of axial/radial lead type devices. Two binding
posts separated by an interval of 7 mm ensure optimum
contact of lead components.
Applicable Instrument: 4294A + 42942A, (4287A, 4395A
with Opt. 4395A-010 + 43961A, 4396B with Opt. 4396B-010
+ 43961A, E4991A)**
When used with 16085B: 4263B, 4268A, 4279A*, 4284A*,
4285A, 4288A, E4980A, E4981A
* denotes the instrument is obsolete.
** applicable in a limited frequency range.
Frequency: DC to 250 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Compensation and Measurement: Open and short
compensations are recommended in combination with the
electrical length compensation before measurement. The
fixture’s electrical length must be entered into the electri-
cal length compensation function of the measurement
instrument first. Open compensation is performed by not
having the binding posts be connected to anything. Short
compensation is performed by using the shorting plate.
After performing open and short compensations in combi-
nation with the electrical length compensation, the DUT
is inserted into the test fixture.
Necessary Accessories:
The shorting plate and the operating note are not fur-
nished as standard. The shorting plate should be ordered
separately. The operating note is downloadable from
Agilent Website (www.agilent.com). Search the web
using the keyword “16093A”.
Description P/N Qty.
Shorting Plate 16092-08010 1
Post spacing