User`s guide

Chapter 3 3-31
Functional Testing
Frequency Response
Table 3-3 Frequency Response Report
Model Center
Freq
Analyzer
Amplitude
Power
Meter
Amplitude
Flatness
Error
Flatness
Error Test
Limits
(dB)
All Models 50 MHz ± 1.5
100 MHz ± 1.5
750 MHz ± 1.5
1250 MHz ± 1.5
1500 MHz ± 1.5
E4402B–
E4408B
2000 MHz ± 1.5
2500 MHz ± 1.5
3000 MHz ± 1.5
E4404B –
E4408B
4250 MHz ± 3.0
5750 MHz ± 3.0
6700 MHz ± 3.0
E4405B –
E4408B
8000 MHz ± 3.5
9000 MHz ± 3.5
10000 MHz ± 3.5
11000 MHz ± 3.5
13200 MHz ± 3.5
E4407B &
E4408B
14000 MHz ± 4.0
19000 MHz ± 4.0
24000 MHz ± 4.0
26500 MHz ± 4.0