Frequency Tracking Calibration with Agilent 11974 Series Preselected
Harmonic Mixers
Using External Millimeter Mixers
(Option AYZ)
Step 8. Change the analyzer center frequency to the value indicated in
Table 10-2 and again adjust the corresponding potentiometer on the
rear panel of the Agilent 11974 until one or both of the green LEDs are
lit.
Step 9. Repeat steps 6 and 7 until the green LEDs are lit at both frequencies.
11974V 50.0 GHz “50.0 GHz Adjust”
Table 10-2 Stop Frequency Preselector Adjustment
Mixer
Agilent P/N
Analyzer Center
Frequency
Potentiometer
11974A 40.0 GHz “40.0 GHz Adjust”
11974Q 50.0 GHz “50.0 GHz Adjust”
11974U 60.0 GHz “60.0 GHz Adjust”
11974V 75.0 GHz “75.0 GHz Adjust”
Table 10-1 Start Frequency Preselector Adjustment (Continued)
Mixer
Agilent P/N
Analyzer Center
Frequency
Potentiometer