Specifications

Chapter 2 13
Measuring Multiple Signals
Comparing Signals on the Same Screen Using Marker Delta
Measuring Multiple Signals
Step 4. Place a marker at the highest peak on the display (10 MHz):
Press
Peak Search.
The
Next Pk Right and Next Pk Left softkeys are available to move the
marker from peak to peak. The marker should be on the 10 MHz
reference signal:
Step 5. Anchor the first marker and activate a second marker:
Press
Marker, Delta.
The label on the first marker now reads 1R, indicating that it is the
reference point.
Step 6. Move the second marker to another signal peak using the front-panel
knob or by using the
Peak Search key:
Press
Peak Search, Next Peak or
Press
Peak Search, Next Pk Right or Next Pk Left.
The amplitude and frequency difference between the markers is
displayed in the active function block. For ESA see the left side of
Figure 2-2 and the right side for PSA.
Figure 2-2 Using the Delta Marker Function (ESA left, PSA right)
NOTE The resolution of the marker readings can be increased by turning on
the frequency count function.