Specifications

Chapter 12 107
Using Segmented Sweep (ESA-E Series Spectrum Analyzers)
Using Segmented Sweep With Limit Lines
Using Segmented Sweep (ESA-E
Series Spectrum Analyzers)
Step 10. Enter points 2 through 11 from Table 12-1:
Step 11. Turn the limits on:
Press
Return, Limit (On).
Step 12. Turn the limit test on to determine if the fundamental signal and
harmonics pass:
Press
Test (On).
Figure 12-3 Segmented Sweep with Limit Line Test
This procedure can verify compliance with phase noise and harmonics
specifications. You can save this instrument state and limit lines in a
setup type file for future applications (
File, Save, Type, Setup,
Save Now). Replace the 10 MHz reference signal of the analyzer with
the device under test (DUT) to create a similar measurement.
Table 12-1 Limit Line Data
Point Frequency Amplitude Connected
1 9.75 MHz
− 80 dBm Yes
29.8 MHz
75 dBm Yes
39.9 MHz
70 dBm Yes
4 9.95 MHz 6 dBm Yes
5 10.05 MHz
6 dBm Yes
610.1 MHz
70 dBm Yes
710.2 MHz
75 dBm Yes
8 10.25 MHz 80 dBm Yes
9 19.95 MHz
30 dBm Yes
10 29.95 MHz
15 dBm Yes
11 30.05 MHz 15 dBm Yes