Specifications

Chapter 12 105
Using Segmented Sweep (ESA-E Series Spectrum Analyzers)
Measuring Harmonics Using Segmented Sweep
Using Segmented Sweep (ESA-E
Series Spectrum Analyzers)
Press Center Freq, 150, MHz.
Press
Span, 100, kHz.
Press
Video BW, 30, Hz.
Notice that the total sweep time for the 3 segments is only 8.13 seconds
(plus a minimal time for transitions between segments). This is
considerably shorter than the 193.3 seconds required to view the same
signals without segmented sweep.
Figure 12-2 Reducing Sweep Time with Segmented Sweep