User`s guide
20
General Information
Specifications
Table 1-6 11970V Specifications
Description Specifications Supplemental Information
RF Frequency Range 50 to 75 GHz
LO Harmonic Number
1
1. When used with the N9030A, harmonic 12 and harmonic 14 used are dependent on the start and
stop frequencies. Harmonic 12 is used from 50 to 66 GHz and harmonic 14 is used above
66 GHz.
14
Odd Order Mixing Product
Suppression
>15 dB (nominal)
LO Input Frequency Range 3.55 to 5.33 GHz
Gain Compression Level (<1 dB) –3 dBm (nominal)
Maximum Conversion Loss 40 dB
Calibration Accuracy
2
2. Calibration accuracy is defined as the difference between the actual mixer conversion loss and
the calibration values provided with the mixer.
±2.2 dB with LO amplitude range of
14.5 dBm to 16.0 dB (nominal)
±3.2 dB with LO amplitude range of
16 dBm to 18.0 dB (nominal)
Spectrum Analyzer Absolute
Amplitude Accuracy (using
conversion loss data with 14.5 to
16.0 dBm LO)
±3.4 dB + frequency response of the
mixer (nominal)
Frequency response (with LO input
power 14.5 to 16.0 dBm)
±2.1 dB
Frequency response (with LO input
power 14.0 to 18.0 dBm)
±2.8 dB
Displayed Noise Level
RBW = 1 kHz RBW
Sample detector
Comversion loss corrections On
LO Power 14.5 to 16 dBm
–92 dBm (nominal)
For N9030A:
50 to 66 GHz: –96 dBm (nominal)
>66to 75 GHz: –94 dBm (nominal)
Input SWR <2.6:1 (nominal)