User`s guide

11
General Information
Introduction
11970 Series Harmonic Mixers
Figure 1-1
Options
Option 009, shown in Figure 1-2 on page 13, is a Mixer Connection Kit. It includes three low-loss SMA
cables (Part Number 5061-5458), one hex-head balldriver (Part Number 8710-1539) for tightening the
waveguide connector screws, and one 5/16-inch open-end wrench (Part Number 8710-0510) for use on
the SMA connectors.
Spectrum Analyzer Retrofit Requirements
N9030A Signal Analyzer
A Diplexer is required since the N9030A has a single SMA external mixer port, and the 11970
series mixers have separate LO and IF ports. The recommended diplexer can be purchased from
Agilent as p/n N9029AE13 or from OML Inc. as p/n DPL.313B.
Requires Option EXM
To retrofit Option EXM:
Software revision must be A.08.01 or later
USB storage device (>2GB free memory)
The N9030A does not support the 11970K (LO range limitation).
N9030A Frequency Option
Serial Number Prefix
<MY/US/SG5138 >
MY/US/SG5138
550, 544, or 543 N9030AK-EXM
N9030AK-EXM
503, 508, 513, or 526 N9030AK-HL6