Technical data

16 Chapter 1
Installation and Setup
Safety Information
Installation and Setup
Safety Considerations For This Analyzer
WARNING This is a Safety Class 1 Product (provided with a protective
earth ground incorporated in the power cord). The mains plug
shall be inserted only in a socket outlet provided with a
protected earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
WARNING Failure to ground the analyzer properly when using the
external power supply can result in personal injury. Before
turning on the analyzer, you must connect its protective earth
terminals to the protective conductor of the main power cable.
Only insert the main power cable plug into a socket outlet that
has a protective earth contact. DO NOT defeat the
earth-grounding protection by using an extension cable, power
cable, or autotransformer without a protective ground
conductor.
WARNING If this analyzer is to be energized via an autotransformer (for
voltage reduction), make sure the common terminal is
connected to the earth terminal of the power source.
WARNING If this product is not used as specified, the protection provided
by the equipment could be impaired. This product must be used
only in a normal condition (in which all means for protection
are intact).
WARNING Whenever it is likely that the protection has been impaired, the
analyzer must be made inoperative and be secured against any
unintended operation.
WARNING To prevent electrical shock, disconnect the Agilent
Technologies spectrum analyzer from mains before cleaning.
Use a dry cloth or one slightly dampened with water to clean
the external case parts. Do not attempt to clean internally.
WARNING When operating from an AC power source, always use the
three-prong ac power cord supplied with this product. Failure
to ensure adequate earth grounding by not using this cord may
cause personal injury and/or product damage.
This product is designed for use in Installation Category II and
Pollution Degree 3 per IEC 61010 and IEC 60664 respectively.