11713A Operating and Service Manual 8-3
Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation
• Before connecting any coaxial cable to an instrument connector for
the first time each day, momentarily ground the center and outer
conductor of the cable.
• Handle all PC board assemblies and electronic components only at
static-safe work stations.
• Store or transport PC board assemblies and electronic components
only in static-shielding containers.
• PC board assembly edge-connector contacts may be cleaned by using
a lintfree cloth with a solution of 80% electronics-grade isopropyl
alcohol and 20% deionized water. This should be performed at a
static-safe work station.
Table 8-1 Static-Safe ESD Accessories
Order the above by calling any Agilent Technologies Sales and Service
Office.
Agilent Part Number Description
9300-0797 Set includes:
3M static control mat 0.6 m x 1.2 m (2 ft x 4 ft) and 4.6 m
(115 ft) ground wire (The wrist-strap and wrist-strap cord
are not included. They must be ordered separately)
9300-0865 Ground wire, 4.6 m (15 ft)
9300-0980 Wrist-strap cord 1.5 m (5 ft)
9300-1367 Wrist-strap, color black, stainless steel, without cord, has
four adjustable links and a 7-mm post-type connection.
9300-1308 ESD heel-strap (reusable 6 to 12 months)