Service manual

11713A Operating and Service Manual 2-5
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
GPIB Addressing
8. Use the five-bit address switches located on the rear panel of the
instrument to change the GPIB address.
The 11713A has a factory preset address of decimal 28. The
switch labeled with a one is the least significant bit.
There are 32 possible addresses. Addresses 0 and 31 are
typically reserved for GPIB functions and should not be used.
IEE-488.1 limits the number of addressable elements
(instruments) to 16.
Figure 2-2 GPIB Address Selection
GPIB Address
Table 2-1 ASCII Address Codes to Decimal Equivalents
ASCII Address
Codes - LISTEN
Decimal
Equivalents
ASCII Address
Codes - LISTEN
Decimal
Equivalents
SP 00 0 16
!011 17
“022 18
#033 19
$044 20
%05 5 21
&06 6 22
‘077 23
(088 24
)099 25
*10 : 26
+11 ; 27
,12< 28
-13= 29
.14> 30
/15? 31