User`s guide
Agilent 1100 Series Valves 57
Specifications 6
* The G1159A 1100 Series 6 positions selection valve can be used at flow rates up to 100 ml/min, but
without valve switching. In most cases e.g. column selection the valve switches during the postrun
or prerun, when the flow can be reduced.
Tabl e 28 G1160A Agilent 1100 Series 12 position/ 13 port selection valve
Liquid contacts: Stainless steel and PEEK
Port size: Accepts 10-32 male threaded fittings
Flow passage diameters: 1.0-mm (0.040”)
Volume in flow passage: Stator (includes stator face seal) 6.4 µl/hole, rotor seal 4.0
µl/groove
Maximum pressure: 21 MPa (207 bar, 3000 psi)
Recommended flow range: 0.2 - 100 ml/min (at high pressures, after the pump)
0.2 - 10 ml/min (at low pressures, in front of the pump)
Tabl e 29 G1162A 1100 Series 2 position/ 6 port micro valve
Liquid contacts: DuraLife processed stainless steel (stator) and vespel (rotor seal)
Port size: Accepts M4 male threaded fittings
Flow passage diameters: 0.20 mm (0.008")
Volume in flow passages: 70 nl port to port
Maximum pressure: 41 MPa (408 bar, 6000 psi)
Recommended flow range: 0.1 - 100 µl
Tabl e 30 G1163A 1100 Series 2 position/ 10 port micro valve
Liquid contacts: DuraLife processed stainless steel (stator) and vespel (rotor seal)
Port size: Accepts M4 male threaded fittings
Flow passage diameters: 0.20 mm (0.008")