28 Agilent 1100 Series Valves
3 Maintenance, Repair, and Troubleshooting
Maintenance and Repair
Resetting the EMF counter
1 Switch to the Diagnostics View of the ChemStation.
2 Select Valve from the Maintenance menu.
3 In the upcoming dialog box click on Reset Counter (see also Figure 18 on
page 32).
The procedure to reset the EMF counter with the 1100 Series handheld
controller is described on page 17.