Technical data

82 1100 Series DAD and MWD Reference Manual
3 Troubleshooting and Test Functions
Heater Failed
Every time the deuterium lamp or the tungsten lamp is switched on or off a
heater self-test is performed. If the test fails an error event is created. As a
result the temperature control is switched off.
Probable Causes
Defective connector or cable.
Defective heater.
Suggested Actions
Ensure the connector to the detector main board is seated firmly.
Exchange heater assembly.