Technical data

1100 Series DAD and MWD Reference Manual 153
Repairing the Detector 4
The figure below shows the required air flow
direction.
8 Insert the fan assembly into its location.
9 Reconnect the connectors to the main board (fan J16),
temperature sensor (J18) and heater (J17), see figure
below for location.
Next Steps:
Reinstall the top foam section, top cover and front cover,
see “Replacing the Foam and the Top Cover” on
page 175.
Turn on the detector and prove, that the fan rotates
smoothly without any scraping noise. Otherwise repeat
step 5 and step 6 on page 152.
Replace the detector into the stack.
Reconnect cables, capillaries and turn on detector
AIRFLOW
G1315/65A
G1315/65B