Technical data

118 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
13 Fix the new heat exchanger to the clamp unit.
Next steps:
Reconnect the capillaries, see “Exchanging a Flow
Cell” on page 105.
Perform a leak test.
Insert the flow cell.
Replace the front cover.
Perform a wavelength calibration, see “Wavelength
Recalibration” on page 84 or holmium test, see
“Holmium Oxide Test” on page 86 to check the correct
positioning of the flow cell.