User`s guide

3Capturing Data
90 Agilent 1000 Series Oscilloscopes User’s Guide
To set up pattern triggers
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select
“Pattern”.
4 Then, either push the entry knob or press Mode again.
5 Press Channel and continue pressing the softkey or turn the entry
knob to select the input channel whose value you want to specify:
6 Press Code and continue pressing the softkey or turn the entry knob
to select the value for the selected channel:
7 Repeat steps 5 through 6 to select the desired values for all input
channels.
The oscilloscope triggers when all the values in the pattern occur at the
same time.
CH1 - CH4 The oscilloscope input channel.
EXT The external trigger input.
EXT/5 The (5:1) attenuated external trigger input.
H Logical high value.
L Logical low value.
X Don’t care value.
Rising edge.
Falling edge.