User`s guide
Displaying Data 2
Agilent 1000 Series Oscilloscopes User’s Guide 57
Using Reference Waveforms
You can save a reference waveform to an internal, nonvolatile memory
location and then display it on the oscilloscope along with other captured
waveforms.
You can also export/import reference waveforms to/from an external USB
drive when it is connected to the front panel USB host port.
Reference waveforms are displayed (that is, turned on/off) just like other
waveforms (see page 43).
To save a reference waveform
1 Before saving a waveform as a reference, set the waveform’s scale and
position as desired.
These settings will become the reference waveform’s defaults.
2 Press [REF].
3 In the REF menu, press Source, and continue pressing the softkey or
turn the entry knob to select the waveform you want to save.
4 Press Location to choose “Internal”.
5 Press Save.
To export or import reference waveforms
To export or import from external storage (when a USB drive is
connected to the front panel USB host port):
1 Press [REF].
2 If exporting a waveform, in the REF menu, press Source, and continue
pressing the softkey or turn the entry knob to select the waveform
you want to export.
NOTE
The reference waveform function is not available in X-Y mode.