Measurement Category
The 1000 Series oscilloscope is intended to be used for measurements in
Measurement Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not
directly connected to MAINS. Examples are measurements on circuits not
derived from MAINS, and specially protected (internal) MAINS derived
circuits. In the latter case, transient stresses are variable; for that reason,
the transient withstand capability of the equipment is made known to the
user.
Measurement category II is for measurements performed on circuits
directly connected to the low voltage installation. Examples are
measurements on household appliances, portable tools and similar
equipment.
Measurement category III is for measurements performed in the building
installation. Examples are measurements on distribution boards,
circuit-breakers, wiring, including cables, bus- bars, junction boxes,
switches, socket- outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example, stationary motors
with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of
the low-voltage installation. Examples are electricity meters and
measurements on primary overcurrent protection devices and ripple
control units.
Transient Withstand Capability
CAUTION
Maximum input voltage for analog inputs:
• CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
• with N2862A/N2863A 10:1 probe: CAT I 600 V (DC + peak AC)