User`s guide
4 Making Measurements
110 Agilent 1000 Series Oscilloscopes User’s Guide
To use manually adjustable cursors
You can set up two parallel, manually adjustable cursors to make
amplitude (vertical) or time (horizontal) measurements on a selected
waveform.
1 Press [Cursors].
2 In the Cursors menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select
“Manual”.
4 Press Ty pe to toggle between:
5 Press Source, and continue pressing the softkey or turn the entry
knob to select the channel or math waveform on which to make the
measurement.
6 To adjust the cursors:
• Press CurA and turn the entry knob to adjust the “A” cursor.
• Press CurB and turn the entry knob to adjust the “B” cursor.
• Press CurA and CurB and turn the entry knob to adjust both
cursors at the same time.
The cursor values displayed are:
• CurA.
• CurB.
• X or Y — difference between CurA and CurB values.
• 1/X — when measuring time parameters, shows the frequency
associated with the time period.
Time To use cursors to measure time parameters.
Amplitude To use cursors to measure amplitude parameters.