Technical data

1Service
38 Agilent 1000A Series Oscilloscopes Service Guide
Table 15 Time Scale Accuracy
Limits Measured time scale error (ppm) Pass/Fail
±50 ppm from 0 °C to 30 °C
±50 ppm + 2 ppm per °C from 30 °C to 45 °C
+ 5 ppm * (years since manufacture)
Table 16 Trigger Sensitivity
Test Limits Ch1 Ch2 C3 Ch4
5 mV/div: 1.5 division at maximum bandwidth
5 mV/div: 1 division at 10 MHz
< 5 mV/div: 1.5 division at 20 MHz
< 5 mV/div: 1 division at 10 MHz